LV power supplys

Table of SCS controlled LV power supply nodes

Name IP Type LAN-Version max. Voltage Task/Detector Nominal Voltage Comment
  hadps001     192.168.100.31              
hadps002 192.168.100.32  
hadps003 192.168.100.33   GEN50-30-LAN     1U1K-IP:4.5-2.0.OCT08-CVP18P1   50 V   MDC/TOF/Central   +48V  
hadps004 192.168.100.34  
hadps005 192.168.100.35 GEN8-180-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 8 V MDC +3.8V  
hadps006 192.168.100.36  
hadps007 192.168.100.37  
hadps008 192.168.100.38 GEN8-180-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 8 V MDC +3V  
hadps009 192.168.100.39 GEN50-30-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 50 V RPC +48V  
hadps010 192.168.100.40 GEN50-30-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 50 V RPC +48V  
hadps011 192.168.100.41 GEN50-30-LAN 1U1K:5.0.1-LAN:2.0 50 V RPC/SHOWER/FW +48V  
hadps012 192.168.100.42  
hadps013 192.168.100.43 GEN8-180-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 8 V MDC +6V  
hadps014 192.168.100.44 GEN8-180-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 8 V MDC -3V  
hadps015 192.168.100.45 GEN8-180-LAN 1U1K-IP:4.5-2.0.OCT08-CVP18P1 8 V MDC +1.8V  
hadps016 192.168.100.46 GEN50-30-LAN 1U1K:5.0.1-LAN:2.0 50 V RICH +48V  


Manuals

TDK Lambda Genesys


Tips, Tricks, and HowTo

LAN reset/defaults

  • LAN reset (p17) LAN manual:
    LAN reset (p17) LAN manual
  • LAN defaults:
    LAN defaults

How to set up parallel operation of TDK Lambda power supplies

    Only power supplies of the same model ( e.g 8V-150A or 50V-30A ) can be connected to increase the output current. Up to 3 slave supplies can be connected to one master supply. The control connection is made only via the master supply.

    • prepare master and slave setup
      1. switch the power on
      2. press the FINE button for three seconds
      3. rotate the CURRENT encoder for the master to the appropriate setting according to the total number of supplies (H2 for 1 slave, H3 for 2 slaves, H4 for 3 slaves)
      4. for the slave rotate the encoder to S (for slave)
      5. press FINE button again and release or wait at least 5 seconds
      6. switch off the power to store the setting in NVRAM
      7. Front panel display after master/slave set up:
        Front panel display after master/slave set up

    • make local programming connections
      1. (on the slave(s) (?)) set switch SW1 position 2 to ON position
      2. cable pin 25 on connector J1 on master to pin J1-10 on all slaves (vertically)
      3. connect pin J1-8 to pin J1-12 on all slaves (locally)
      4. Switch SW1 settings and local programming cable:
        Switch SW1 settings and local programming cable

    • make solid connections between all +V0 terminals and all -V0 terminals and connect to load
      • The good connection between the -V0 terminals is essential for safe operation.
      • Solid output power connections:
        Solid output power connections:

    • To restore to original setting (single supply)
      • remove local programming cable
      • on master and slaves
        1. press FINE for 3 seconds, rotate CURRENT encoder to H1
        2. press FINE button again and release or wait at least 5 seconds
        3. switch off to store in NVRAM
      • (on the slave(s) (?)) set SW1 position 2 to OFF position

    -- BurkhardKolb - 31 Mar 2010



-- PeterZumbruch - 23 Mar 2022
ISorted ascending Attachment Action Size Date Who Comment
LANdefault.pngpng LANdefault.png manage 44 K 2012-10-01 - 17:35 PeterZumbruch LAN defaults
LANreset.pngpng LANreset.png manage 75 K 2012-10-01 - 17:30 PeterZumbruch LAN reset (p17) LAN manual
Topic revision: r9 - 2022-03-23, PeterZumbruch
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki Send feedback | Imprint | Privacy Policy (in German)