You are here: Hades Wiki>HODO Web>WebHome (30 Jan 2015, JochenMarkert)EditAttach

Hodoscope detectors

Three pieces of Hodoscope detectors for beam position monitoring are available. Two of them are identical with scintillator rod dimensions 100 x 10 x 5 mm, the third one has smaller rods (100 x 6.5 x 6.5 mm). All three detectors show significant light attenuation inside the scintillators and had to be repaired in Optical laboratories of the Charles University in Prague, Czech Republic.


Hodoscope publication in NIM: Pion_beam-hodoscope.pdf

Photomultipliers datasheet: Photomultiplier_tube_R3478.pdf

Hodoscope drawings

created in SimpleGeo (

Hodoscope drawings ppt: Hodoscope_drawings.pptx

Hodoscope drawings pdf: Hodoscope_drawings.pdf

Source file Hodoscope_drawings_source.dat

Material file materials.db

High Voltage for PMTs

How to set HV

HV values on Hodoscope nr.2: HV_for_hodoscope.pdf


vncviewer hadesdaq01:1 is main DAQ VNC

vncviewer hadesdaq01:2 is used for HODO/START

Thresholds on PaDiWa Amps

Setting the thresholds on PaDiWa front-end boards Thresholds

TDC calibration of TRB3 TRB3calibration

Look-up table (24. 4. 2014): Lookup_table_start_category.pdf

Attenuation before repair

Attenutaion length before repair was 4 - 7 cm depanding on the rod and method of measurement (cosmic muons, gamma or beta radiation).

Hodoscope nr.3 - measurement with 90Sr radioactive source: Hodoskop3_attenuation.pdf

After repair

Attenuation length after exchange of scintillator rods is cca 24 cm.

Short report about attenuation measurement after repair: after_repair.pdf

Beam results

Pulse amplitudes on protons measured with osciloscope:

Pulse amplitudes

"Heat map" with well focused beam: Beam profile on hodoscope (Y axis):

Heat map Profile

QA over web browser:

open web browser from any machine in GSI network: hades33/mon/

or from outside:

or from WiFi:

OR via SOCKS proxy setting:

1. open a ssh connection and activate 'dynamic port forwarding'
ssh -D 1080

2. Setup SOCKS proxy in Firefox
open tab 'Network'
click on Connection 'Settings...'
activate 'Manual Proxy Configuration'
SOCKS Host: localhost Port: 1080
(ensure that all other Proxy fields are blank (or 0))

How to analyze the data (may 14 test):

QA was moved to

0.) go to folder /analysis/tof/May14

1.) to find last 20 hld files type:


ls -ltr /store/**/*.hld(.om[0,20]L+33)

2.) run placed in the /makeDST/ folder

3.) use analysis macro from /makeDST_ilse to analyse the files (./analysisDST name_of_the_file_from_step_1 number_of_events)

Beamtime info


exchange of 2 modified Padiwas for lower rods



LookUp table and CalParameters for START and HODO: calibpar_notshifted.txt

HODO Web Utilities

Topic attachments
I Attachment Action Size Date Who Comment
HV_for_hodoscope.pdfpdf HV_for_hodoscope.pdf manage 251.1 K 05 May 2014 - 10:43 OndrejSvoboda High Voltage for Hodoscope nr.2
Hodoscope.datdat Hodoscope.dat manage 331.0 K 30 Sep 2014 - 06:49 OndrejSvoboda Hodoscope drawings source file
Hodoscope_drawings.pdfpdf Hodoscope_drawings.pdf manage 1315.0 K 30 Sep 2014 - 06:33 OndrejSvoboda Hodoscope drawings pdf
Hodoscope_drawings.pptxpptx Hodoscope_drawings.pptx manage 3799.5 K 30 Sep 2014 - 06:35 OndrejSvoboda Hodoscope drawings pptx
Hodoskop3_attenuation.pdfpdf Hodoskop3_attenuation.pdf manage 531.4 K 05 May 2014 - 12:23 OndrejSvoboda Attenuation measurement with Hodoscope nr.3 and 90Sr source
Lookup_table_start_cathegory.txttxt Lookup_table_start_cathegory.txt manage 48.6 K 06 May 2014 - 09:24 OndrejSvoboda Look-up table for start cathegory
Photomultiplier_tube_R3478.pdfpdf Photomultiplier_tube_R3478.pdf manage 215.7 K 05 May 2014 - 10:34 OndrejSvoboda Photomultiplier tube from hodoscope
Pion_beam-hodoscope.pdfpdf Pion_beam-hodoscope.pdf manage 440.0 K 05 May 2014 - 10:30 OndrejSvoboda Design and commissioning of the GSI pion beam - hodoscope description
after_repair.pdfpdf after_repair.pdf manage 260.0 K 06 May 2014 - 07:25 OndrejSvoboda 90Sr measurement of hodoscope nr. 2 after repair
calibpar_notshifted.txttxt calibpar_notshifted.txt manage 6.1 K 19 Aug 2014 - 19:33 LukasChlad LookUp plus CalPars (before shift)
hodo_pulse_amplitude.jpgjpg hodo_pulse_amplitude.jpg manage 145.1 K 05 May 2014 - 12:03 OndrejSvoboda Pulse amplitudes from Hodoscope nr.2 stripe 0, protons 2 GeV
hodoscope.jpgjpg hodoscope.jpg manage 283.6 K 06 May 2014 - 06:51 OndrejSvoboda hodoscope layout
hodoscope_heatmap.jpgjpg hodoscope_heatmap.jpg manage 7.0 K 05 May 2014 - 12:31 OndrejSvoboda Hodoscope heatmap with well focused beam
hodoscope_profile.jpgjpg hodoscope_profile.jpg manage 3.3 K 05 May 2014 - 12:37 OndrejSvoboda Beam profile in Y axis seen in Hodoscope
materials.dbdb materials.db manage 3.0 K 30 Sep 2014 - 06:51 OndrejSvoboda Hodoscope drawings material file
Topic revision: r17 - 30 Jan 2015, JochenMarkert
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Hades Wiki? Send feedback